Product Overview
Harmonic Filter Function (Available in the Yokogawa 30032A)
1. What is a Harmonic?
Harmonic refers to sinusoidal quantity having a frequency that is an integral multiple of the fundamental frequency (for example, the commercial frequency). When a harmonic is superimposed on the fundamental frequency, the waveform is distorted.
2. Why is it Necessary to Accurately Measure the Leakage Current of the Fundamental Frequency (Commercial Frequency)?
One problem when measuring the leakage current to check the isolation of electrical circuits in power distribution equipment is that the electrical
isolation cannot be correctly understood due to the influence of a harmonic current. That is, the leakage current flowing from the electrical circuit
to ground is very small so that, in order to check the isolation of electrical circuits by means of the leakage current, it is necessary to remove the
harmonic component of the leak current and measure only the current of the fundamental frequency (commercial frequency).
3. The Yokogawa 30032A Employs a Harmonic Filter
Conventional leakage clamp-on testers could not sufficiently remove harmonic current components so measured leakage current values were often larger than the specified value due to the influence of a harmonic current. In this case, retesting with an insulation tester was required, resulting in increased effort and cost for the test. Under these circumstances, Yokogawa Meters & Instruments Corporation has developed the leakage clamp-on tester 30032A, which employs a high-performance harmonic filter that can accurately measure just the fundamental frequency component of the leakage current.
Product Specifications
Description | Leakage Current Mini Clamp-on; 3/30mA/30/60 Amp |
---|---|
Method of detection | Mean Value |
Display | (Bar-graph display: 32 segments) |
Range Switching | Auto-range |
Operating temperature and humidity | 0 - 50°, 80% RH or less (no condensation) |
Influence of external magnetic field | 0.0005% typical 1 (on current value of adjacent cable) |
Influence of conductor position | Within accuracy |
Safety Standard | Approved for conformity to EN 61010-1,EN 61010-2-032 |
Circuit voltage | 300 Vrms or less |
Withstanding voltage | 3.7 kV AC for one minute |
Power Supply | One CR2032 (Coin-shaped) 3 V battery |
Battery Life | Approx. 90 hours (when continuously used) |
Auto power-off | Approx. 10 minutes |
Diameter of measurable conductor | 40 at maximum mm |
Dimensions | Approx. 70 (W) x 176 (H) x 25 (D) mm |
Weight | Approx. 200 g |
Accessories | Instruction manual, battery, Soft case |
AC Current | 3/30 mA, 30/60A |
Frequency Characteristics | 50/60 Hz |
Accuracy (50/60Hz) | 1.0 + 5 (0~30mA) |
Measurement Cycle | 2 times/second (Digital display), 12 times/second (Bar-graph display) |
Temperature Coefficient | 0.05% of range /°C or less (within the ranges of 0-18° and 28-50° for measurement of 0-50 A) |
Note |
|
At 23°C±5°C,75%RH or less
Accuracy: ±( % rdg + dgt) = ±(% readout + value of least significant digit)
AC current | 3/30mA, 30/60A | |
---|---|---|
Frequency characteristics | 50/60Hz | |
Accuracy | 50/60Hz | 1.0 + 5(0~30mA) |
1.0 + 5(0~50A) | ||
5.0 + 5 (50-60A) |
AC current measurement | |||
Filter function OFF | |||
Accuracy: ± (% of reading + digits) | |||
Range | Resolution | Accuracy | Maximum |
3mA | 0.001mA | 0.010 < I ≤32.70mA: | 3.270mA |
30mA | 0.01mA | 32.70mA | |
30A | 0.01A | 0.05 < I ≤50.0A:1.0%+5 | 32.70A |
60A | 0.1A | 60.6A | |
Filter function ON | |||
3mA | 0.001mA | 0.010 < I ≤32.70mA: | 3.270mA |
30mA | 0.01mA | 32.70mA | |
30A | 0.01A | 0.05 < I ≤50.0A:1.5%+5 | 32.70A |
60A | 0.1A | 60.6A | |
Note: Input current of 2nd-order and higher harmonics |
Key Features
◆ Leakage current measurement range from 3mA to 60A
◆ Diameter of measurable conductor: 40 mm max.
◆ Easier and clear display
Accessoires Included
◆ User’s manual
◆ Batteries
◆ Soft carrying case (RB057)
Additional Information
Weight | 200 g |
---|---|
Dimensions | 70 (W) x 176 (H) x 25 (D) mm |
※ 자세한 사항은 'DataSheet'나 'Quick Start Guide'를 참조해 주세요.
상품이 장바구니에 담겼습니다.